『Address of Shaanbei Quality Flow Meter Manufacturer』Related information(clamp on meter|electromagnetic meter|venturi meterrotameter|orifice meter|ultrasonic flow meter|mass flow meter|coriolis mass flow meter|coriolis flow meter|magnetic flow meter|magmeter flow meter|magflow flow meter|mag meter flow meter|electromagnetic flow meter|vortex flow meter|turbine flow meter|thermal mass flow meter|thermal flow meter|rotameter flow meter)

1. Compact Thermal Mass Flow Meter Installation and Setting Guide
Compact Thermal Mass Flow Meter Installation and Setting Guide
1. Installation Precautions Flow Direction: When installing a flowmeter, be sure to observe the arrow indicating the flow direction on the housing to ensure that the flow direction of air/gas is consistent with the arrow on the housing. This is the foundation for ensuring accurate measurement of flow meters. Requirements for straight pipe section: Although the flowmeter is integrated with a flow regulator and does not require an additional straight pipe section, in order to achieve the best measurement effect, it is recommended to use 3-5 times the inner diameter of the fiber large straight pipe at the inlet. Installation position: The flowmeter can be installed on horizontal or vertical pipelines, and the specific installation position should be determined according to the actual working conditions and pipeline layout. Process connection: For DN
8, DN1
5, DN20, DN25, and DN32 pipelines, G internal thread connection is used. Ensure that the connection is well sealed to prevent gas leakage.
Secondly, the electrical connection flowmeter is equipped with two wiring plugs "A" and "B". By default, it comes with a 5-meter cable with an M8 female connector on one end and an open wire on the other end. For flow meters, a single connecting wire can make them work properly. If it is necessary to separate the output pulse or the power supply and signal separately, two connecting wires need to be ordered.
3. Relevant Settings Download APP: After normal installation and wiring, a dedicated APP needs to be downloaded (Android phones can click on the download link, I


Current position > Product detail